N2O Wastewater Sensor 3.0
The N2O Wastewater Sensor is the digital sensor body that holds the replaceable Sensor Head. It delivers stable measurements across aerobic, anoxic, and anaerobic zones, so you can follow both where N2O is produced and where it is reduced. It is built to handle the harsh wastewater conditions where foam, condensation, and fouling are part of normal operation.
The fully digital sensor stores calibration and sensor-specific data on the sensor, so you can move it or replace its Sensor Head without losing your setup or history. The sensor calculates your N2O emissions and sends the results to the Operator Console. You connect your airflow signal to the Connector Unit, so you get direct emission results without additional programming or calculations. Read more...
The N2O Wastewater System 3.0 consists of
Measure N2O directly in your process
Calibrate at the tank
You only need to calibrate every second month, and it takes 5–10 minutes. You can do it right next to the tank with a bucket of 4 l tap water and the N2O Calibration Kit. The sensor has a built-in temperature sensor, so you just follow the LED guidance from the N2O Wastewater Connector Unit.
Smart sensor operation
The sensor continuously monitors its own status and alerts you when it is time to calibrate or replace the Sensor Head. This helps you stay on top of maintenance without manual checks and supports continuous measurements. Because calibration and sensor data are stored on the sensor, you keep your calibration, history and setup when you move the sensor or replace its head.
The sensor keeps more than two years of sensor and emission data and a logbook of more than one year.
Part of a flexible monitoring system
To measure N2O, you use the N2O Wastewater Sensor with three more components: a Sensor Head, a Connector Unit and an Operator Console. You need one Sensor Head per sensor, and Sensor Heads are ordered separately. Each Connector Unit takes up to 2 sensors, and one Operator Console takes up to 16 sensors. You integrate the data directly with your PLC or SCADA, with no subscriptions or mandatory cloud platforms.
Learn more about the N2O Wastewater System.
| Feature | Description |
|---|---|
|
Feature
Material
|
Description
Anodized aluminium alloy | Robust design in aluminium alloy casing (6063-T6) and black POM acetyl copolymer
|
|
Feature
Dimensions & weight
|
Description
60mm x 390mm (D x L), 1.2 kg
|
|
Feature
Mounting
|
Description
Chain Loop and stainless steel chain kit
|
|
Feature
Power Supply
|
Description
24V DC from N2O Wastewater Connector Unit
|
|
Feature
Communication
|
Description
ModbusRTU (Slave)
|
|
Feature
N2O signal
|
Description
1 x temp. compensated N2O value (N2O-N [mg/L] or ppm)
|
|
Feature
Temperature
|
Description
1 x temperature (°C or °F)
|
|
Feature
Emission Rates
|
Description
1 x Emission calculations (N2O-N [mg/m3/d]) with dynamic input parameters | Emissions from anaerobic, anoxic and aerobic zones
|
|
Feature
Emission Parameters
|
Description
kLaN₂O, vgas
|
|
Feature
Sensor Health
|
Description
Sensor & calibration status: needs calibration, needs replacement, warnings & errors | Diagnostic data: Raw and calibration signals
|
|
Feature
Sensor Head
|
Description
Required, ordered separately: standard, medium or high measuring range, for 0–27 °C or 27–40 °C.
|
Quick Enquiry
Related Products
Measure N2O directly in the liquid phase to optimize your wastewater treatment process and reduce your emissions.
Operate and calibrate your sensors where they are installed.
Collect and manage your N2O measurement data in one place.
Measure dissolved N2O with fast response time in the biological tanks of wastewater treatment plants
Collect data from up to two sensors and transmit directly to your PLC or SCADA system
The replaceable Clark-type measuring element of the N2O Wastewater Sensor
Calibrate your nitrous oxide sensor every second month
Free placement of your N2O Wastewater Controller
Robust sensor mounting for fast flowing water
Compare N2O Wastewater System generations