Video Guides
The video guides below will show you the only maintenance needed for the N2O Wastewater System, calibration and sensor head exchange.
Click the gear wheel in the video player, then Subtitles, then Auto-translate to choose your preferred language for subtitles.
Please be aware that the subtitles are machine-translated and may contain minor errors. If you have any questions, we are always happy to help by chat, phone, or e-mail.
Calibration and Sensor Head Exchange
Two point calibration of N2O Wastewater Sensor
This video will show each step to perform for a two point calibration on our N2O Wastewater Sensor.
Exchange the N2O Wastewater Sensor Head
Learn how you exchange the N2O Wastewater Sensor Head in this 2-minute video.
Zero point calibration of N2O Wastewater Sensor
Watch this video for a complete, step-by-step guide on how to perform a zero point calibration on our N2O Wastewater Sensor.
Troubleshooting: Remove an Air Bubble
Depending on conditions during shipping and storage, microbubbles invisible to the eye can be stuck in the tip of the sensor. These will cause a faulty signal – often a sensor that is fixed at maximum or minimum value. To dislodge such air bubbles, shake the sensor as shown in the video.
Please note: A large air bubble is always found in the tip – it is part of the sensor construction.